USER GROUP PROCEDURES ( 7411B ) 7411B PROCEDURE ============================================================================== INSTRUMENT: HP 410C DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 100% NUMBER OF TESTS: 43 NUMBER OF LINES: 161 CONFIGURATION: Fluke 5100B w/Wideband Fluke 5205A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Ensure that the Test Instrument has Warmed up. 1.001 DISP Turn the Test Instrument Power Off. 1.002 DISP Zero the Meter Movement using the Center Screw . 1.003 ASK- F 1.004 DISP Switch the Test Instrument On. 1.005 OPBR Does the Test Instrument have a rear ZERO ADJ? 1.006 JMPF 1.101 1.007 DISP Set the FUNCTION switch to +DCV 1.008 DISP Set the RANGE to .015V. 1.009 DISP Short the DCV lead (grey) and COM lead (black) together. 1.009 DISP Adjust the ZERO ADJ on the rear of the Test Instrument 1.009 DISP for a 0 reading on the meter. 1.009 DISP Switch between +DCV and -DCV and ensure the meter 1.009 DISP remains zeroed. 1.010 DISP Remove the short. 1.011 DISP Set the FUNCTION to ACV 1.012 DISP Set the RANGE to .5V. 1.013 DISP Short the AC PROBE Tip and Ground lead together. 1.014 DISP Using the AC ZERO control in the center of the FUNCTION 1.014 DISP switch, ZERO the METER. 1.015 DISP Remove the short. 1.016 DISP Connect the AC PROBE end to the 'Binding Posts to Double 1.016 DISP Banana Plug Adapter' red terminal by placing it in the 1.016 DISP small side hole and tightening the head of the adapter. 1.016 DISP This probe should be attached from the right hand side 1.016 DISP facing the red plastic head at the top. 1.016 DISP Connect the GROUND of the Probe to the Black terminal. 1.017 HEAD {Volts AC Tests} 1.018 DISP The following tests use the AC PROBE and Adapter for 1.018 DISP connection to the Test Interface Panel. 1.019 ASK- R 1.020 5100 0.5 0.5V 3/ 400H 2W 2.001 DISP Set the RANGE to 1.5V 2.002 DISP Short the AC PROBE Tip and Ground Lead Together. 2.002 DISP ZERO the meter using the AC ZERO control. 2.002 DISP Remove the short. 2.003 5100 1.5 1.5V 3/ 400H 2W 3.001 DISP Set the RANGE to 5V 3.002 DISP Short the AC PROBE Tip and Ground Lead Together. 3.002 DISP ZERO the meter using the AC ZERO control. 3.002 DISP Remove the short. 3.003 5100 5 5V 3/ 400H 2W 4.001 DISP Set the RANGE to 15V 4.002 DISP Short the AC PROBE and ground lead together. 4.002 DISP ZERO the meter using the AC ZERO control. 4.002 DISP Remove the short. 4.003 5100 15 15V 3/ 400H 2W 5.001 ASK+ R 5.002 5100 50 50V 3/ 400H 2W 6.001 5100 150 150V 3/ 400H 2W 7.001 5100 500 300V 3/ 400H 2W 8.001 DISP Disconnect the AC PROBE and Adapter from the TI Panel. 8.002 DISP Connect the 'Binding Posts to BNC Plug Adapter' on to 8.002 DISP the Adapter now connected to the AC PROBE. 8.003 DISP Connect the 50ohm Feedthrough to the BNC Plug connected 8.003 DISP to the AC PROBE. 8.004 HEAD {Frequency Response Tests} 8.005 DISP The following tests use the AC PROBE and Adapters for 8.005 DISP connection to the Test Interface Panel. 8.006 DISP Set the RANGE to .5V 8.007 DISP Short the AC PROBE Tip and Ground Lead Together. 8.007 DISP ZERO the meter using the AC ZERO control. 8.007 DISP Remove the short. 8.008 ASK- R 8.009 WIDE 0.5 0.5V 3/ 400H N 2W 8.010 MEMC 0.5 .5V 30% 400H_Reference 9.001 WIDE 0.5 0.5V 10/ 20H C 2W 10.001 WIDE 0.5 0.5V 2/ 100H C 2W 11.001 WIDE 0.5 0.5V 2/ 1MH C 2W 12.001 WIDE 0.5 0.5V 2/ 10MH C 2W 13.001 ASK+ R 13.002 DISP Remove the connections from the TI Panel. 13.002 DISP Remove the Adapters from the AC PROBE, and in a similar 13.002 DISP manner, connect the DCV (grey) and COM (black) leads to 13.002 DISP the 'Binding Posts to Double Banana Adapter' only. 13.003 HEAD {Volts DC and Linearity Tests} 13.004 DISP The DCV, COM leads and Adapter will be used for the 13.004 DISP following connections to the Test Interface Panel. 13.005 DISP Set the FUNCTION switch to +DCV. 13.006 5100 0.015 0.015V 2/ 2W 14.001 5100 0.05 0.05V 2/ 2W 15.001 5100 0.15 0.15V 2/ 2W 16.001 5100 0.5 0.5V 2/ 2W 17.001 5100 1.5 1.5V 2/ 2W 18.001 5100 5 1V 2/ 2W 19.001 5100 5 2.5V 2/ 2W 20.001 5100 5 5V 2/ 2W 21.001 5100 15 15V 2/ 2W 22.001 5100 50 50V 2/ 2W 23.001 5100 150 150V 2/ 2W 24.001 5100 500 500V 2/ 2W 25.001 5205 1500 1000V 2/ 2W 26.001 DISP Set the FUNCTION switch to -DCV. 26.002 5205 1500 -1000V 2/ 2W 27.001 DISP Remove the connections from the TI Panel. 27.001 DISP Remove the DCV lead from the Adapter and replace it 27.001 DISP with the DCA/OHMS lead (red). 27.002 HEAD {Resistance Tests} 27.003 DISP The DCA/OHMS, COM leads and Adapter will be used for 27.003 DISP the following connections to the Test Interface Panel. 27.004 DISP Set the FUNCTION switch to OHMS. 27.005 ASK- R 27.006 DISP Select the RX10M range. 27.007 DISP Ensure that the Test Instrument leads are not touching. 27.007 DISP Adjust the Rear OHMS infinity ADJ for full-scale on the 27.007 DISP meter. 27.008 OPBR Does the meter have a rear panel ZERO ADJ control? 27.009 JMPF 27.101 27.010 DISP Short the DCA/OHMS and COM LEADS TOGETHER at the Adapter 27.010 DISP Adjust the ZERO ADJUST on the rear of the Test 27.010 DISP Instrument for a 0 reading on the meter. 27.010 DISP Remove the short. 27.011 DISP Set range to {RX10} 27.012 RESF 10Z 0.5U 2W 28.001 DISP Set Range to {Rx100} 28.002 RESF 100Z 5U 2W 29.001 DISP Set Range to {Rx1K} 29.002 RESF 1KZ 0.05U 2W 30.001 DISP Set Range to {Rx10K} 30.002 RESF 10KZ 0.5U 2W 31.001 DISP Set Range to {Rx100K}. 31.002 RESF 100KZ 5U 2W 32.001 DISP Set Range to {Rx1M}. 32.002 RESF 1MZ 0.05U 2W 33.001 DISP Set Range to {Rx10M}. 33.002 RESF 10MZ 0.5U 2W 34.001 ASK+ R 34.002 HEAD {Current DC Tests} 34.003 DISP Set the FUNCTION switch to +DCA. 34.004 5100 15 15uA 3/ 2W 35.001 5100 50 50uA 3/ 2W 36.001 5100 150 150uA 3/ 2W 37.001 5100 0.5 0.5mA 3/ 2W 38.001 5100 1.5 1.5mA 3/ 2W 39.001 5100 5 5mA 3/ 2W 40.001 5100 15 15mA 3/ 2W 41.001 5100 50 50mA 3/ 2W 42.001 5100 150 150mA 3/ 2W 43.001 DISP Set FUNCTION to -DCA. 43.002 5100 150 -150mA 3/ 2W END